K-Assay
Immunoassay Reagents
For Use on Automated Chemistry Analyzers
 
Serum FDP
Cat. No.: KAI-320
Type: Reagent
Method: Immunoturbidimetric
Format: Liquid
Linearity: 0.2 - 80 µg/mL (or value of highest calibrator)
Sample: Serum


KIT COMPOSITION
Cat. No. KAI-320
R1:   2 X 12 mL buffer
R2:   1 X 5 mL antibody

Tests / Kit: 125 (Hitachi 917)
(Can Vary Based On Analyzer Model)
    
KAI-320
Click Image To Enlarge       

Click here for an example package insert.

Click here for the Safety Data Sheets (SDS):
Reagent-1 (R-1)
Reagent-2 (R-2)

Click here for information about recommended calibrators and controls.


INTENDED USE
For the quantitative determination of fibrin / fibrinogen degradation products in human serum.
FOR RESEARCH USE ONLY IN THE U.S.
NOT FOR USE IN DIAGNOSTIC PROCEDURES IN THE U.S.

REAGENT PREPARATION
The reagents are ready to use and do not require reconstitution.  Before use, gently invert Reagent 2 at least once a week.

STORAGE AND HANDLING
All reagents should be stored refrigerated (2-8°C).  Return all reagents to 2-8°C promptly after use.  Unopened reagents can be used for 18 months from the date of manufacture, as indicated by the expiration date on the package and bottle labels.  Opened reagents can be used for one month if stored at 2-8°C.

PERFORMANCE
Precision
The following within run precision results were abtained on a Roche/Hitachi analyzer with pooled human serum or urine:

  Sample 1 Sample 2
N 20 20
Mean 3.97 µg/mL 14.42 µg/mL
S.D. 0.0979 0.1725
C.V. 2.47 % 1.20 %

Accuracy / Correlation
A comparison of the K-ASSAY ®   Serum FDP reagent and another company's Serum FDP reagent was performed with the following results on serum samples:

Linear Regression: y = 0.9697x + 0.2924
r = 0.9990
n = 62
x = another company's Serum FDP assay
y = K-ASSAY ®   Serum FDP Assay

Assay Range
0.2 - 80 µg/mL (or value of highest calibrator)

INTERFERENCE
Bilirubin, Conjugated:     No interference ≤ 20 mg/dL
Bilirubin, Unconjugated:     No interference ≤ 20 mg/dL
Chyle (Formazine Turbidity)    No interference ≤ 3,000 FTU
Hemoglobin:     No interference ≤ 500 mg/dL
Rheumatoid Factor:     No interference ≤ 570 IU/mL