K-Assay
 
Clinical Diagnostics - Immunoassay Reagents
For Use on Automated Chemistry Analyzers
 
IgA
Immunoglobulin A
Cat. No.: KAI-013
Type: Reagent
Method: Immunoturbidimetric
Format: Liquid
Linearity: 20 - 800 mg/dL
Sample: Serum


KIT COMPOSITION
R1:   4 X 20 mL buffer
R2:   2 X 10 mL antibody

Tests / Kit:   285 (Hitachi 717)
(Can Vary Based On Analyzer Model)
    
KAI-013
Click Image To Enlarge       

Click here for an example package insert.

Click here for the Safety Data Sheets (SDS):
Reagent-1 (R-1)
Reagent-2 (R-2)

Click here for information about recommended calibrators and controls.


INTENDED USE
For the quantitative determination of human IgA in serum by imunoturbidimetric assay.   Measurement of IgA aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.
FOR IN VITRO DIAGNOSTIC USE.

REAGENT PREPARATION
The reagents are ready to use and do not require reconstitution.

STORAGE AND HANDLING
All reagents should be stored refrigerated (2-8°C).  Return all reagents to 2-8°C promptly after use.  Unopened reagents can be used for one year from the date of manufacture, as indicated by the expiration date on the package and bottle labels.  Opened reagents can be used for one month if stored at 2-8°C.

PERFORMANCE
Precision

Precision - Within Run
Sample I Sample II Sample III
N = 20 N = 20 N = 20
Mean = 150 Mean = 296 Mean = 443
S.D. = 1.12 S.D. = 2.06 S.D. = 4.52
C.V. = 0.7% C.V. = 0.7% C.V. = 1.02%

Precision - Between Runs
Sample I Sample II Sample III
N = 7 N = 7 N = 7
Mean = 92 Mean = 269 Mean = 356
S.D. = 1.98 S.D. = 6.91 S.D. = 8.47
C.V. = 2.2% C.V. = 2.6% C.V. = 2.4%

Accuracy / Correlation
A comparison of the K-ASSAY ®   IgA assay and an INCSTAR IgA Test Kit was performed on a Hitachi 704 automated analyzer and a COBAS Mira.  The test results provided the following data:

y = 1.0188x - 6.73
r = 0.9899
n = 45
x = INCSTAR IgA Test Kit
y = K-ASSAY ®   IgA assay

x min = 39 y min = 48
  max = 780   max = 769
  mean = 243   mean = 241

Assay Range
20 - 800 mg/dL

INTERFERENCE
Bilirubin :     No interference up to 20 mg/dL
Hemoglobin :     No interference up to 500 mg/dL
Intralipid :     No interference up to 500 mg/dL