K-Assay
 
Immunoassay Reagents
For Use on Automated Chemistry Analyzers
 
Apo E / Apolipoprotein E
Cat. No.: KAI-007
Type: Reagent
Method: Immunoturbidimetric
Format: Liquid
Linearity: 1 - 15 mg/dL
Sample: Serum, Plasma


KIT COMPOSITION
R1:   3 X 20 mL buffer
R2:   1 X 20 mL antibody

Tests / Kit:   200 (Hitachi 717)
(Can Vary Based On Analyzer Model)
    
KAI-007
Click Image To Enlarge       

Click here for an example package insert.

Click here for the Safety Data Sheets (SDS):
Reagent-1 (R-1)
Reagent-2 (R-2)

Click here for information about recommended calibrators and controls.


INTENDED USE
For the quantitative determination of human Apolipoprotein E (Apo E) in serum and plasma by immunoturbidimetric assay.
FOR RESEARCH USE ONLY IN THE U.S.
NOT FOR DIAGNOSTIC PROCEDURES IN THE U.S.

REAGENT PREPARATION
The reagents are ready to use and do not require reconstitution.

STORAGE AND HANDLING
All reagents should be stored refrigerated (2-8°C).  Return all reagents to 2-8°C promptly after use.  Unopened reagents can be used for one year from the date of manufacture, as indicated by the expiration date on the package and bottle labels.  Opened reagents can be used for one month if stored at 2-8°C.

PERFORMANCE
Precision
When a sample that has an apolipoprotein E concentration around 4 mg/dL is assayed 20 times (within-run), the absorbance C.V. is below 5%.

Assay Range
1 - 15 mg/dL

INTERFERENCE
Bilirubin :     No interference up to 20 mg/dL
Hemoglobin :     No interference up to 500 mg/dL
Intralipid :     No interference up to 2,000 mg/dL