K Logo Kamiya Biomedical Company
K-ASSAY
Since 1983
K-Assay
 
Beta-2 Microglobulin / Ferritin -
Serum / Plasma Control
Cat. No.: K283C-4M
Type: Control


KIT COMPOSITION
Cat. No. K283C-4M
Level 1 Human Serum 2 x 3 mL, lyophilized
Level 2 Human Serum 2 x 3 mL, lyophilized
    
K283C-4M
Click Image To Enlarge       

Click here for an example package insert.

Click here for the Safety Data Sheet (SDS).


INTENDED USE
The K-ASSAY ®   Beta-2 Microglobulin / Ferritin - Serum / Plasma Control is intended for use as a consistent test sample of known concentration for monitoring the performance of the K-ASSAY ®  Beta-2 Microglobulin and K-ASSAY ®  Ferritin immunoturbidimetric assays with serum or plasma samples.
FOR IN VITRO DIAGNOSTIC USE.

CONTROL PREPARATION
Add exactly 3 mL of purified water to the vial containing the lyophilized serum to dissolve it. Let stand for 20 minutes. Invert gently until fully dissolved.

STORAGE AND HANDLING
Store controls at 2-8°C.  Return all controls to 2-8° promptly after use.   Unopened controls can be used until the expiration date on the package and bottle labels.   Opened, reconstituted controls can be used for 2 weeks if stored at 2-8°C.

EXPECTED VALUES
The following values are given as an example.  The expected values for the K-ASSAY ®   Beta-2 Microglobulin / Ferritin - Serum / Plasma Control are continually being revised through ongoing quality assurance.   As a result, the expected values may change from lot to lot.  Please refer to the package insert and/or vial labels for each lot for the expected values.

 REAGENT   SAMPLE TYPE   UNIT   LEVEL 1   LEVEL 2 
 MEAN   RANGE   MEAN   RANGE 
 K-ASSAY ®  Beta-2 Microglobulin   Serum / Plasma   mg / L   1.82   1.55 -
2.09 
 7.88   7.09 -
8.67 
 K-ASSAY ®  Ferritin   Serum / Plasma   ng / mL   105   89 -
121 
 396   337 -
455